Parks Medical Electronics, Inc.
   
 

NOTICE - PARKS EQUIPMENT IS CLASS II - NON-INVASIVE. Click Here

 

INSTRUCTIONAL
Programming Your 3100-B IR Remote

REMOTE “LEARN MODE”

When replacing a remote, the 3100-B must learn to recognize the new remote. This was done at the factory for the original 3100-B and remote pair.

1. Turn OFF the 3100-B.

2. Hold the VPR button on the front panel of the 3100-B down, and turn ON the 3100-B. The lower front panel LEDs will flash continuously.

3. Press and hold the YELLOW button on the remote. When the LEDs on the lower front panel of the 3100-B stop flashing, the process is complete.

3100-B remote

 

 

TROUBLESHOOTING
Troubleshooting your 3100-B IR remote

If you are having trouble with your 3100-B IR remote control, here are schematics to help you isolate the problem.

3100-B charger base schematics 297-0314-00.00
3100-B charger base parts list MBOM7-0314-00.00
3100-B charger base parts layout REFD0314-01.00

 

3100-B IR remote schematics 299-0310-02.00
3100-B IR remote parts list MBOM9-0310-02.00
3100-B IR remote parts layout REFD0290-01.00